				AlarmA new generation of area alarm fromShire
				Controls Ltd
Features:-Compact size: 244mm wide, 178mm high & 45mm deep.Low cost.
 Simple connection: Just 3 cores connect the alarm to the end
				of line components, regardless of the number of gases.
 Simple Maintenance: All connections (including mains) are plug
				& socket. The alarm components can be changed in about 1.5
				minutes.
 Versatile: Several SAX-6 alarms can be
				connected to the same end of line components using just 3 cores. The SAX-6 can
				transmit either a common alarm or all alarm conditions, using the MEDCON data
				protocol, for dispaly on the SDX-15 plant alarm system, and can also recive
				data from the SDX-15 system.
 2 year warranty: No
				parts or labour charges for repairs to alarms which fail due to faulty parts or
				manufacture.
 10 year comapatibility warranty: Install the SAX-6 with the
				confidence that equipment will be available to modify or extend the system for
				at least 10 years.
 Complies with: HTM2022, BS EN 737
				part 3 (the new European standard) and BS EN 4754
 
 The SAX-6 area alarm is used to monitor pipeline
				pressure within a ward, theatre etc. The pressure is monitored by pressure
				switches in the pipeline downstream from the last AVSU. Pressure is monitored
				for both high and low pressure (low only for vacuum). The integrity of the
				cabling between the pressure switches and the alarm panel is monitored, and a
				fault on this cable will result in a system fault alarm, with all affected
				alarm conditions going into alarm condition.
 
 The monitoring of the
				cable requires a termination box to be mounted as close as practical to the
				pressure switches. As an optional extra, these termination boxes can be
				supplied with the cables for connection to the pressure switches already
				fitted, allowing significant savings in on-site work.
 To use more than one
				SAX-6 alarm panel with one set of pressure switches, simply connect all SAX-6
				alarm panels to the 3 core screened cable (minimum 0.5mm) from the termination
				box (either at the termination box itself, or at another SAX-6).
 External Indicators:-SAX-6 has 3 LED indicators marked P, M & D on the front panel which indicate the following:
 The P LED represents pressure switch inputs.
 
 If the LED is off the panel has not been set up to detect pressure switch inputs via the end of line transmitter board, (this can be overcome on the
jumper setting shown in our manual).
 
 If the LED is flashing, the SAX panel has been set up to receive pressure switch inputs, but has not been connected (or connected in error) to the end
of line transmitter board.
 
 If the LED is on, this indicates that the SAX panel and transmitter board are set up and communicating correctly.
 
 The M LED represents use with our MEDCON protocol (as with our SDX system).
 
 If the LED is off the panel has not been set up to work with our SDX system and NO plug-in upgrade card has been installed.
 
 If the LED is flashing, then a plug-in card has been installed to enable use with our SDX system, but the 2-core data buss used for such systems has not
been connected (or is connected incorrectly).
 
 If the LED is on, this indicates that the SAX panel is working in conjunction with the SDX network and is capable of transmitting the local
alarm conditions onto the SDX system.
 
 The D LED is reserved for units with the main FPGA IC programmed to D type. The LED is either on or off and is designed to show at a glance whether a unit incorporates the D type chip without having to open each unit up.
 
 The purpose of the chip is to enable 2no 3 gas panels to provide status to 1no 6 gas central unit using standard boards and cabling (or indeed 2no 2 gas board providing status to a 4 gas unit).
 
 
 Installation requirements:A 230 vac supply, fused at 3 amps, fed from the
				essential supply
 A 3 core screened cable (minimum 0.5mm) from the alarm
				panel to the termination board
 Termination Boxes
 When ordering, please
				specify the following:- Flush or surface mounting.Number of gases
 If possible, the hospital in which the alarm is to be installed.